This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author's copyright. In most cases, these works may not be reposted without the explicit permission of the copyright holder.

Search for Publication


Year(s) from:  to 
Author:
Keywords (separated by spaces):

Thermal expansion imaging for monitoring lesion depth using M-mode ultrasound during cardiac RF ablation: in vitro study

Peter Baki, Sergio J Sanabria, Gabor Kosa, Gabor Székely, Orcun Goksel
International Journal of Computer Assisted Radiology and Surgery
Vol. 10, No. 6, pp. 681-693, June 2015

Abstract

PURPOSE: We demonstrate a novel method for automatic direct lesion depth (LD) tracking during coagulation from time series of a single A-mode ultrasound (US) transducer custom fit at the tip of a RFA catheter. This method is named thermal expansion imaging (TEI). METHODS: A total of 35 porcine myocardium samples were ablated (LD 0.5-5 mm) while acquiring US, electrical impedance (EI) and contact force (CF) data. US images are generated in real time in terms of echo intensity (M-mode) and phase (TEI). For TEI, displacements between US time series are estimated with time-domain cross-correlation. A modified least squares strain estimation with temporal and depth smoothing reveals a thermal expansion boundary (TEB)-negative zero-crossing of temporal strain-which is associated to the coagulated tissue front. RESULTS: M-mode does not reliably delineate RFA lesions. TEI images reveal a traceable TEB with RMSE [Formula: see text] 0.50 mm and [Formula: see text] with respect to visual observations. The conventional technique, EI, shows lower [Formula: see text] and [Formula: see text]200 % variations with CF. The discontinuous time progression of the TEB is qualitatively associated to tissue heterogeneity and CF variations, which are directly traceable with TEI. The speed of sound, measured in function of tissue temperature, increases up to a plateau at 55 [Formula: see text], which does not explain the observed strain bands in the TEB. CONCLUSIONS: TEI successfully tracks LD in in vitro experiments based on a single US transducer and is robust to catheter/tissue contact, ablation time and even tissue heterogeneity. The presence of a TEB suggests thermal expansion as the main strain mechanism during coagulation, accompanied by compression of the adjacent non-ablated tissue. The isolation of thermally induced displacements from in vivo motion is a matter of future research. TEI is potentially applicable to other treatments such as percutaneous RFA of liver and high-intensity focused ultrasound.


Link to publisher's page
@Article{eth_biwi_01203,
  author = {Peter Baki and Sergio J Sanabria and Gabor Kosa and Gabor Székely and Orcun Goksel},
  title = {Thermal expansion imaging for monitoring lesion depth using M-mode ultrasound during cardiac RF ablation: in vitro study},
  journal = {International Journal of Computer Assisted Radiology and Surgery},
  year = {2015},
  month = {June},
  pages = {681-693},
  volume = {10},
  number = {6},
  keywords = {Ultrasound · Strain imaging · Lesion control · Radio-frequency ablation}
}