This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author's copyright. In most cases, these works may not be reposted without the explicit permission of the copyright holder.

Search for Publication


Year(s) from:  to 
Author:
Keywords (separated by spaces):

Submillimeter Measurement of Cup Migration in Clinical Standard Radiographs

K. Burckhardt, G. Székely, H. Nötzli, J. Hodler, and C. Gerber
IEEE Transactions on Medical Imaging
Vol. 24, pp. 676-688, May 2005

Abstract

Assessing the displacement of bony implants is an important topic in arthroplasty, particularly in total hip replacement (THR). The observation of the migration is supposed to provide an insight into the fixation of the implant. Diagnostic standard radiographs of the pelvis are an advantageous data source for this purpose. The previous methods based on these images, however, lack of a thorough consideration of their projective nature. They hence do not reach the desired precision, which should lie in the sub-millimeter range to allow a detection of migration in the first one or two years after implantation. The aim of the work presented here was, therefore, a method for measuring the distance of the artificial hip socket to the bone with an error of less than 0.5 mm. The approach has been on the one hand to define the bone-cup distance measured in the radiograph so that the variability of the intrinsic and extrinsic parameters at exposure has a minimal impact. On the other, specialized matching techniques are applied in order to optimize the localization of the necessary bony landmarks and the cup in the X-ray image. The coordinates of the bony landmarks are determined by means of a template matching algorithm. The position of the implant is estimated by intensity-based registration using the cup's CAD-model. The method was validated theoretically, experimentally, and clinically. In the clinical radiographs, the standard deviation of the migration measurements resulted to be 0.28 mm when using only natural bony landmarks. The implantation of a bony marker was found to increase the precision to a standard deviation of 0.20 mm. The inter-observer variability in the two cases was estimated to be 0.11 mm and 0.04 mm.


Download in pdf format
@Article{eth_biwi_00393,
  author = {K. Burckhardt and G. Székely and H. N\"otzli and J. Hodler and and C. Gerber},
  title = {Submillimeter Measurement of Cup Migration in Clinical Standard Radiographs},
  journal = {IEEE Transactions on Medical Imaging},
  year = {2005},
  month = {May},
  pages = {676-688},
  volume = {24},
  number = {},
  keywords = {CAD-model, hip implant, image matching, migration measurement, radiography}
}